1

Direction

What do you know, and what do you want computed?
2

The vial

Required. Not inferred, defaulted or suggested, including for named reagents. Activity content (IU, U) accepts scientific notation, e.g. 1e6.

What the stated content is the content of

A portion taken from a bulk solid is weighed and entered as the content itself: "weighed, from bulk".

Conjugate state
Carrier protein

Reagent, carrier, excipient and salts together, where the certificate states it. Used for the volume the dry solids occupy. Without it, displacement is not computable.

3

Target and volume

Which volume is meant

Choose a direction first.

4

Capability and vessel

suggested, not chosen: 2 µL. Replace it with your syringe's or pipette's value.

Molecular weight from the Molarity Converter

optional, for a molar target or molar form

Nothing entered into this tool's fields is stored or transmitted. Reloading the page clears every field.

About this calculator

What it does

This tool answers one bench question in either direction: how much diluent to add to a vial to reach a target concentration, or what concentration a stated volume gives. It also records what the stated content is the content of, where the figure came from, whether carrier is present, the conjugate state, and whether the volume taken up by the dry solids was accounted for.

The arithmetic of reconstitution is simple. What is not trivial is what the number on the vial is the amount of, what else is in the vial, and how much volume the dry solids take up, and nothing in the resulting solution reveals any of them. This tool asks for each and puts it in the result.

How to use it

  1. Choose the direction. "Target → volumes" if you know the concentration you want; "volume → concentration" if the datasheet says "reconstitute in 1 mL".
  2. Enter the stated vial content and its unit from the label or certificate. The tool never infers or suggests it. Mass units (µg, mg, g) and activity units (IU, U) are different things and are never converted into each other.
  3. Say what the content is the content of. "The reagent alone" if the figure is the protein or molecule of interest only. "The total vial contents" only if the figure is the whole dry mass in the vial: reagent, carrier protein, excipient and buffer salts together. A figure such as "1 mg total protein" counts the reagent and carrier but not sugars or salts, so it is neither; choose "not recorded" for it and keep the datasheet wording in your notes. If you cannot tell, choose "not recorded" and the result says so.
  4. Say where the figure came from: a lot-specific certificate of analysis (the content measured for that lot), the vial label or datasheet (the nominal fill), or weighed from bulk by you. The tool records this; it does not correct for fill tolerance.
  5. Declare the conjugate state and carrier. If the molecule carries a label or payload, say whether the stated mass is of the protein alone or of the whole conjugate. Say whether the stock contains carrier protein, because a protein assay of it will then report the carrier too.
  6. Enter the total solids mass if the certificate gives it (reagent, carrier, excipient and salts together). It lets the tool compute the volume the dry solids occupy. Without it, only the reagent's own volume is corrected and the concentration is reported as "at most".
  7. Confirm the whole vial is reconstituted. A fraction of a cake, or a vial already reconstituted, is outside this tool.
  8. Enter the target concentration or the volume, and say which volume you mean: the diluent you draw up and inject, or the final volume the vial will contain.
  9. Declare your minimum reliable transfer volume (replace the 2 µL suggestion with your syringe's or pipette's value) and, optionally, the vial's working capacity.
  10. For a molar target or molar form, the molecular weight must come from the Molarity Converter as a transport link, which carries its source, its mass-basis declaration and its flags. The Molarity Converter does not produce that link yet, so molar targets cannot be computed here for now; every mass-based result is unaffected.
  11. Read the result and every flag, then choose "Copy for notebook". The copy records the declarations, the constants applied, the flags and the engine version.

How it works

Dry solute occupies volume. Adding 1 mL of diluent to 100 mg of protein does not give 1 mL of solution, and there is no mark on a sealed vial to bring it to. Where the total solids mass is known, the displacement is D = solids × v̄ and the diluent to add is smaller than the final volume by D. Where it is not, mass-based reagent-alone content is corrected for the reagent's own volume only and its concentration is reported as at most; activity content and content basis not recorded are not corrected.

The diluent volume, the only number anyone physically delivers, is always rounded from its own unrounded value. The concentration reported is the one obtained when that displayed volume is delivered, Ca = content ÷ (Vd′ + D), so it can differ from the target by up to 0.5 %. It is shown beside the target, not in place of it.

Constants and assumptions

NameValueWhy
Partial specific volume, v̄0.73 mL/gAverage for globular proteins, range ≈ 0.70–0.75: Harpaz, Gerstein & Chothia 1994, Structure 2:641; Perkins 1986, Eur J Biochem 157:169. Scoped to protein-dominated solids. A constant of the tool, not a user input.
Error if the true v̄ is lowerε = (0.73 − v̄true) × total-solids concentrationApplying 0.73 mL/g to solids whose true value is lower over-corrects: the true final volume is smaller than computed by ε of it, and the true concentration is above the reported one by 1 ÷ (1 − ε). At 100 mg/mL of solids with v̄ = 0.3 mL/g: 4.30 % first-order, 4.49 % exact. Below 10 mg/mL, under 0.5 % for any solid with v̄ ≥ 0.3 mL/g. The correction beats no correction where the solids' mass-weighted v̄ exceeds 0.365 mL/g.
Partial specific volumes of common excipientsSucrose, trehalose and mannitol about 0.6 to 0.67 mL/g; sodium chloride about 0.3 mL/g; some salts lowerApproximate values, for orientation only. The calculation never uses them: it applies 0.73 mL/g to all solids. They show why the correction suits protein-, sugar- and sugar-alcohol-rich cakes (above 0.365 mL/g, where correcting beats not correcting) and is not guaranteed for salt-rich ones.
Material displacement threshold0.600 % of the final volume (about 8.2 mg/mL of total solids at 0.73 mL/g)A displacement above this share of the final volume is flagged. 0.6 % is the smallest maximum permissible systematic error in ISO 8655-2:2022, Table 1, which covers single-channel air-displacement pipettes and positive-displacement pipettes with a reusable plunger and capillary. That error is the largest bias in the mean delivered volume that a pipette may have and still pass calibration. The 0.6 % entry is for pipettes whose nominal volume (largest setting) is above 5 mL and up to 20 mL, used at that volume. Every other entry is at least as large: on the same row, 1.2 % at half the nominal volume and 6.0 % at a tenth; for pipettes above 50 µL up to 5 mL, 0.80 % at nominal. So the flag can be raised while the displacement is still smaller than the error permitted for the pipette and setting you are using. That is deliberate: it tells you about the volume of the solids early. These are calibration limits, not the measured error of your own pipette.
Minimum displacement DminDmin = content × v̄; fmin = v̄ × content ÷ F, which is v̄ × target in the target direction: 7.3 % at 100 mg/mL, 0.73 % at 10 mg/mLFor mass-based reagent-alone content, solids ≥ content, so Dmin is a lower bound under the two assumptions below, to within v̄'s scope error. Applied as a partial correction where it is the only displacement defined. Not defined for activity content or content basis not recorded.
Unattainable-volume boundary1 ÷ v̄ ≈ 1370 mg/mL total solidsWhere D (or Dmin) ≥ F the diluent volume is ≤ 0. Evaluated as D ≥ F, or Dmin ≥ F, on computed values. The unattainable-volume rejection is a sign-change guard at about 1370 mg/mL of total solids; it will not fire on any real reconstitution, and its silence is not a plausibility check. It cannot fire with a diluent volume entered.
Displayed-volume resolution0.05–0.5 % of the value, leading-digit dependentA half-step at 3 significant figures is 0.005 on a mantissa between 1.00 and 9.99. The concentration achieved at the displayed diluent volume therefore differs from the target by up to 0.5 %, and the tool reports it rather than bounding it.
Minimum reliable transfer volumeYou declare it; 2 µL is pre-filled as a suggestionThe smallest volume your syringe or pipette delivers reliably. Replace the suggestion with your own value; the notebook copy records whether it was left unchanged.
Vial working capacityYou declare it; optional, no defaultThe largest volume the vial can hold. Without it the tool cannot check the final volume against the vial.
AssumptionApplies to
Volumes are additive: final volume = diluent volume + solids mass × v̄. The correction beats no correction for solids whose mass-weighted v̄ exceeds 0.365 mL/g: protein-, sugar- and polyol-dominated solids. For salt-dominated solids it is not guaranteed.Every result where displacement is computable
Where the total solids mass is not known: mass-based reagent-alone content is corrected for the reagent's own volume only (Dmin), and its concentration is reported as an upper bound. For activity content or content basis not recorded, no correction is applied and no bound is available.Every result where the displacement-not-computable flag is raised
The reagent is a protein, within the scope of v̄. Even then Dmin is a lower bound only to within v̄'s scope error: a reagent whose true v̄ is 0.70 mL/g displaces 0.959 × Dmin: an over-correction of a few tenths of a percent of the volume where the solids are almost entirely reagent, about 0.3 % at 100 mg/mL.Every result reporting Dmin
The solids other than the reagent do not reduce the solution volume. Whether any common lyophilization excipient has a small or negative apparent volume at the relevant concentrations is not established here.Every result reporting Dmin
The solids dissolve completely into the diluent.Every result (see "What it does not do")

Precision, rounding and engine version

Volumes are displayed to 3 significant figures and concentrations to 6; fractions and percentages to 3. Rounding is half away from zero, on the exact binary value, except for a bound: an "at most" concentration is rounded up and an "at least" final volume down, so that the figure shown is itself a bound. The final volume shown is rounded. The concentration is computed from the diluent volume as displayed and the displacement applied, not from the rounded final volume. Every unrounded value is carried in the structured result object. A figure within one unit in the last place of a rounding boundary can fall one step short of the exact bound.

Engine 0.1.5. The version changes whenever the calculation changes, so two results with the same engine version come from the same arithmetic. Result object ligant.bench-tools.result 2.0.0-draft.c7.

Flags

CodeWhat it meansWhat to do
FL-01The concentration is of total solids (reagent, carrier and excipient together), not of the reagent.Do not read it as the reagent concentration. Find the reagent content if you need that.
FL-02It was not recorded whether the stated content is the reagent alone or the total vial contents.Check the datasheet or certificate and run it again with the basis declared.
FL-03The content is the nominal fill from the label or datasheet, not the lot's measured content.The actual fill can differ from the nominal. Use the lot-specific certificate of analysis where one exists.
FL-04The source of the stated content was not recorded.Record where the figure came from before using the stock.
FL-05The total solids mass is not known, so the volume of the solids cannot be computed in full; the concentration is "at most", or uncorrected.Enter the total solids mass if the certificate gives it. Otherwise treat the concentration as an upper bound.
FL-06The final volume exceeds the vial's declared working capacity.The solution will not fit in this vial. Reconstitute to a higher concentration whose final volume fits, then dilute that stock in a separate tube (the Dilution Planner plans this for mass and molar concentrations).
FL-07The diluent volume is below your declared minimum reliable transfer volume.You cannot deliver it reliably. Choose a lower target concentration (or, if you entered a volume, a larger one) so that more diluent is added, use a syringe or pipette that delivers this volume reliably and declare its minimum, or reconstitute a vial with more content.
FL-08The dry solids take up a material fraction of the final volume.Do not add the final volume as diluent. Add the diluent volume shown.
FL-09The molar form is withheld because the declarations do not pair safely with the imported molecular weight.Use the mass-based result, or correct the declarations or the imported molecular weight.
FL-10The imported molecular weight carried flags of its own.Read them; they apply to the molar form.
FL-11The stock contains carrier protein.A total-protein measurement of this stock (A280, BCA, Bradford) includes the carrier. Do not use one to check the reagent's concentration.
FL-12It was not recorded whether carrier is present.Find out before interpreting a protein assay of the stock.
FL-13The concentration is of the protein alone, or of the conjugate, as declared. No degree-of-labelling correction is made.Check that this is the mass your datasheet states.
FL-14It was not recorded whether the stated mass includes a label or payload.Record it before pairing the concentration with a molecular weight.

Why not a spreadsheet?

  • Every assumption is a recorded choice, not a hidden cell. The content basis, its source, the conjugate state, carrier presence, the volume basis and the displacement treatment are declared, and each appears in the copy-out with its effect on the result.
  • The volume of the dry solids is computed, or its absence is flagged. A typical reconstitution sheet divides the content by the diluent volume as if the solids took no space. Here the displacement is applied, bounded, or stated as not computable.
  • Units are carried and checked. Mass against activity, and IU against U, are refused rather than converted; a molar value appears only when the molecular weight and the declarations pair correctly.
  • Rounding happens once, on display. The diluent volume is rounded from its own unrounded value, and the concentration shown is the one you get when you deliver that rounded volume.
  • Flags say in words what to check, and the same inputs always give the same output. The copy-out records the inputs, declarations, constants, flags and engine version.

What it does not do

This tool determines a volume or a concentration. It does not verify what was delivered or whether the solids dissolved. It does not plan dilutions of the stock (Dilution Planner), does not identify the protein, does not supply molecular weights or specific activities, and does not convert between mass and activity. The tool guarantees that the volume arithmetic is correct and that the content basis, its provenance, the conjugate state, carrier presence, the volume basis and the displacement treatment are recorded. It cannot detect:

  1. A stated content that is wrong: mislabelled, or a vial from a different lot than the certificate consulted.
  2. Incomplete dissolution. The tool assumes the solids dissolve fully. A cake that has not fully gone into solution gives a lower concentration than computed, and the solution may look clear.
  3. Loss to the vial and closure. Protein adsorbs to glass and to the stopper, which matters most at the low masses where it is least visible.
  4. Loss during reconstitution: foaming, vortexing, material left on the stopper or wall. The most common loss at the bench, and the tool sees none of it.
  5. Loss of activity before reconstitution: storage, temperature excursion, or age. A correct mass of inactive protein reconstitutes to the computed concentration and does not work.
  6. The wrong diluent. Datasheets frequently specify the diluent: dilute acid for some growth factors, buffer with carrier for others. The tool computes a correct concentration in the wrong solvent and cannot know.
  7. A collapsed, discoloured or partially melted cake, which indicates a compromised vial. The tool sees only the numbers.
  8. Vial vacuum. Many lyophilized vials are stoppered under vacuum, which draws diluent in and can leave delivered volume differing from the volume drawn up.
  9. Whether the built-in partial specific volume applies. The value is for protein-dominated solids; a cake that is mostly excipient has a lower value, and the correction then overshoots by a bounded amount in a stated direction (see the constants table). The tool cannot know the excipient fraction.
  10. Whether a specific activity used elsewhere is the lot's. This tool does not convert between activity and mass, so it never holds one; a user who converts by hand before entering a value brings that risk with them.
  11. Any error in the execution. The tool states the volume; it does not observe what was delivered.
  • Dilution of the resulting stock (Dilution Planner), titration series design (Antibody Titration Planner), multi-component cocktails.
  • Conversion between activity units and mass through a specific activity; molecular-weight lookup or protein identification; buffer or diluent selection, carrier addition, aliquoting or storage guidance; correction for incomplete dissolution or handling loss; reconstitution of a fraction of a vial; degree-of-labelling correction.

The stock envelope carries the full result for use by another tool; the Dilution Planner does not read it, so the notebook copy is the record. An imported object carries a check that detects accidental corruption, truncation, or removal of fields including its flags. This is an integrity check and not an authenticity check: it cannot detect a deliberate edit, because there is no server and no secret, and anyone who edits the object can recompute the check. This tool does not claim otherwise.

Questions and answers

How do I work out how much diluent to add to a lyophilized vial?

Choose the direction: "target → volumes" if you know the concentration you want, "volume → concentration" if the datasheet says "reconstitute in 1 mL". Enter the stated vial content and its unit from the label or certificate; the tool never infers or suggests it. Say what that content is the content of, where the figure came from, the conjugate state, and whether carrier protein is present. Enter the total solids mass if the certificate gives it, so the tool can compute the volume the dry solids occupy. Confirm the whole vial is reconstituted. Enter the target concentration or the volume, and say which volume you mean: the diluent you draw up and inject, or the final volume the vial will contain. Declare your minimum reliable transfer volume, replacing the 2 µL suggestion with your syringe's or pipette's value. Then read the result and every flag, and add the diluent volume shown.

Does the dry powder change the volume when I reconstitute?

Yes. Dry solute occupies volume. Adding 1 mL of diluent to 100 mg of protein gives about 1.07 mL of solution, not 1 mL, and there is no mark on a sealed vial to bring it to. Where the total solids mass is known, the displacement is D = solids × v̄, and the diluent to add is smaller than the final volume by D: add the diluent volume shown, not the final volume. Here v̄ is the partial specific volume, the volume each gram of dissolved solid adds; the tool uses 0.73 mL/g, an average for globular proteins, for all solids. That suits protein-, sugar- and sugar-alcohol-rich cakes and is not guaranteed for salt-rich ones. Where the total solids mass is not known, mass-based reagent-alone content is corrected for the reagent's own volume only and its concentration is reported as "at most"; activity content, and content whose basis was not recorded, are not corrected.

What does the amount printed on the vial mean, and does it matter?

It can be the amount of the reagent alone (the protein or molecule of interest, exclusive of carrier and excipient) or of the total vial contents (the whole dry mass: reagent, carrier protein, excipient and buffer salts together), and nothing in the resulting solution reveals which. It matters: from the total vial contents, the concentration is of total solids, not of the reagent. A figure such as "1 mg total protein" counts the reagent and carrier but not sugars or salts, so it is neither; choose "not recorded" for it and keep the datasheet wording in your notes. If carrier protein is present, a total-protein measurement of the stock (A280, BCA, Bradford) includes the carrier, so do not use one to check the reagent's concentration. The result records the basis, where the figure came from, whether carrier is present and the conjugate state.

Can I reconstitute part of a vial, or a vial I have already reconstituted?

Not with this tool. The whole vial is reconstituted; no fraction of a cake is accepted. A previously reconstituted vial is a liquid stock: plan its dilution in the Dilution Planner, which takes mass and molar concentrations but not activity units (IU/mL, U/mL).

Can it convert micrograms to units?

No conversion is made between mass and activity. Mass units (µg, mg, g) and activity units (IU, U) are different things and are never converted into each other, and IU and U are not equated either. A conversion needs the lot's specific activity, and the tool does not supply specific activities or molecular weights; if you convert by hand before entering a value, check that the specific activity is the lot's.

What can go wrong that the calculator cannot detect?

It states the volume; it does not observe what was delivered. It cannot detect a stated content that is wrong (mislabeled, or from a different lot than the certificate consulted), incomplete dissolution (the solution may look clear), protein lost by adsorption to the glass or stopper, loss during reconstitution (foaming, vortexing, material left on the stopper or wall), loss of activity before reconstitution, the wrong diluent, a collapsed, discolored or partially melted cake, or vial vacuum, which can make the delivered volume differ from the volume drawn up. Nor can it know whether the built-in partial specific volume suits your cake: it is for protein-dominated solids, and for a cake that is mostly excipient the correction overshoots by a bounded amount in a stated direction. It never holds a specific activity, so it cannot check one you used elsewhere.

Is my data uploaded anywhere?

No. Calculations run entirely in your browser, and your inputs are never transmitted, stored, or logged. The page sends an email address only if you choose to submit one in the newsletter signup, and its web analytics counts visits and sees how quickly the page loads, with further analytics only if you allow it in the banner; no analytics ever receives anything you type into this tool.

Is it free?

Yes. It is free to use and open source under the Apache 2.0 licence. Its source code is on GitHub, so you can download it and run it on your own computer.

Privacy

The stock